Search found 2 matches
- Sat Feb 17, 2024 10:09 pm
- Forum: Test Equipment
- Topic: Fluke Friday - Going through the Fluke Collection
- Replies: 158
- Views: 71453
Re: Fluke Friday - Going through the Fluke Collection
Zenith, thank you. I will try to be a bit more active. Probably averaged about one comment/year over at the old place, prior to the TEA blow-up there. Figured I 'd better finally say something here lest I be suspected of engaging in non-consensual relations with valves. Med, the failure in the pots ...
- Fri Feb 16, 2024 6:13 pm
- Forum: Test Equipment
- Topic: Fluke Friday - Going through the Fluke Collection
- Replies: 158
- Views: 71453
Re: Fluke Friday - Going through the Fluke Collection
Fluke Friday service note: If you have a Fluke 8505A or 8506A it might be wise to check the power supply outputs. Each one has three trimpots for adjustment, and each of my three meters had one of those three trimpots fail open, sending the corresponding rail to its maximum output. After the third o...